UV-VIS Spektrofotometre Jenway séria 63
 Zobraziť v plnej veľkosti
|
UV-VIS Spektrofotometre Jenway.
- Wavelength range of 198 to 1000nm
- Absorbance, %T and concentration modes
- Multi-parameter display with wavelength and photometric readouts
- Wide range of sampling accessories
- Domed lid of 6320D accepts tubes up to 105mm tall
- 6305 utilises a high quality pulsed xenon lamp
- Wide range of sampling accessories
The 6305 is high quality, low cost unit for spectrophotometric applications in schools and colleges, as well as in QC and testing procedures for a wide variety of services and industries. The 6320D is based on the 6300 platform and includes a domed lid which accepts taller tubes up to a height of 105mm.
|
6305 |
6315 |
| Wavelength range |
198 to 1000nm |
198 to 1000nm |
| Wavelength resolution |
1nm |
1nm |
| Wavelength accuracy |
±2nm |
±2nm |
| Spectral bandwidth |
8nm, 6nm over UV range |
8nm |
| Transmittance |
0 to 199.9%T |
0 to 199.9%T |
| Absorbance |
–0.300 to 1.999A |
–0.300 to 1.999A |
| Accuracy |
±1%T |
±1%T |
| Resolution |
0.1%T, 0.001A |
0.1%T, 0.001A |
| Stray light |
<0.5%T at 220 and 340nm |
<0.5% at 340nm |
| Noise |
<0.001A at 0A at 400nm |
<0.001A at 0A at 400nm |
| Stability |
<0.002A/h after 30 minutes |
<0.002A/h after 30 minutes |
| Concentration range |
–300 to 1999 |
–300 to 1999 |
| Concentration resolution |
0.1 to 1 |
Selectable 1, 0.1, 0.01 or 0.001 |
| Units |
ppm, mg/l, g/l, M, %, blank |
|
| Factor |
0 to 199.9, 1000 to 9999 |
|
| Light source |
Xenon lamp |
Xenon lamp |
| Outputs |
Analogue and RS232 |
Analogue and RS232 |
| Power |
<50W |
<50W |
| Size (w x d x h) |
365 x 272 x 160mm |
365 x 272 x 160mm |
| Weight |
6kg |
6kg |
| Calibration |
|
Blank with a single standard or factor |
| Quantitation range |
|
–300 to 1999 |
| Quantitation resolution |
|
Selectable 1, 0.1, 0.01 or 0.001 |
| Calibration |
|
Blank with up to 6 standards |
| Curve correction |
|
Linear regression, interpolation or linear regression through zero |
| Kinetics display |
|
Graphical and calculated concentration value |
| Calibration |
|
Against standard and factor |
| Correlation factor |
|
r2 value displayed with result |
| Spectrum range |
|
Any range between 198 and 1000nm |
| Scan interval |
|
1, 2 or 5nm, selectable |
| Analysis |
|
Absorbance and wavelength of peaks and valleys |
| GLP |
|
Real time clock and calendar, Operator ID. Supervisor security (locks all set up parameters). Method display lock |
|
Posledná aktualizácia: Sobota, 19 máj 2012 21:04